Millimar | Module for inductive probes

Millimar N 1702 M-HR

Art. no. 5331125

Module for inductive probes Millimar N 1702 M-HR

  • Flexibly configurable RS-485 bus modules
  • Powerful connection modules for evaluating measuring sensors (inductive)
  • High resolution of measuring signal, 0.01 µm
  • Synchronous data retrieval from multiple connected measuring probes
  • Connection of N 1700 modules via USB interface to the smart, universal evaluation and configuration software Millimar Cockpit
  • Connects all compatible probe types using one module
  • Flexible and modular product combinations to tackle customer-specific measuring tasks
  • Maximum theoretical bus data rate of 4189 values/s (depending on the number of connected channels)

instruction manual

Millimar N 1702 M-HR
Millimar N 1702 M-HR
  • Flexibly configurable RS485 bus modules
  • Powerful connection modules for evaluating measuring sensors (inductive)
  • High resolution of measuring signal, 0.01 µm
  • Synchronous data retrieval from multiple connected measuring probes
  • Connection of N 1700 modules via USB interface to the smart, universal evaluation and configuration software Millimar Cockpit
  • Connects all compatible measuring probe types using one module
  • Flexible and modular product combination to tackle customer-specific measuring tasks
  • Maximum theoretical bus data rate of 4189 values/s (depending on the number of connected channels)
Millimar N 1702 M-HR | Art. no. 5331125
Product type N 1702 M-HR
Resolution µm 0,01
Measuring range, inductive probe ± 200
Probe inputs
Compatibility Mahr , Mahr 1340 , Mahr half-bridge , Mahr LVDT , Mahr VLDT
Configuration Millimar Cockpit Software
Data transmission rate 4189 Hz
Error limit 0,3 % (min. 0,04 μm)
Data interface RS-485
Energy supply + 5V from N 1700 Bus
Current consumption 115 mA
IP protection category IP 42
Working temperature MIN 10 °C
Working temperature MAX 35 °C
Operating temperature Min 0 °C
Operating temperature Max 40 °C
Temperature coefficient 0.02 µm/°C
Product weight 0.27 kg

Variety of applications for our products

Smart and flexible combination of measuring modules and software for solving customer-specific measuring tasks.
Label Art. no. Measuring range from Measuring range up to Product type Distance of lower stop MIN Distance of lower stop MAX Distance of upper stop MAX Lifter / retraction Measuring force (N) Increase in measuring force Sensitivity deviation Repeatability fw (µm) Measuring value hysteresis fu (µm) Linearity deviation within +/- 1,0 mm Linearity deviation within +/- 2,0 mm IP protection category Cable length Working temperature Working temperature MIN Working temperature MAX Operating temperature Min Operating temperature Max Storage temperature MIN Storage temperature MAX Temperature coefficient Compatibility Linearity deviation within +/- 0,5 mm Distance of upper stop MIN Linearity deviation within +/- 0,1 mm Linearity deviation within +/- 0,3 mm
5313400 Inductive probe
1340
5313400 -2 mm 2 mm 1340 -2.2 mm 0 mm 3 mm Vacuum lifter 0,75 N 0,08 N/mm 0.3 % 0.08 µm 0.08 µm 0.15 µm 0.4 µm IP 64 1.5 m 20 °C 10 °C 40 °C -10 °C 80 °C -10 °C 80 °C 0.6 µm/°C Mahr 1340
5313010 Inductive probe
1301
5313010 -1 mm 1 mm 1301 -1.1 mm 0 mm 2.7 mm Cable release 0,75 N +/- 0,15 N 0,4 N/mm 0.3 % 0.1 µm 0.2 µm 2 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.09 µm/°C Mahr LVDT 0.5 µm
5313030 Inductive probe
1303
5313030 -1 mm 1 mm 1303 -1.1 mm 0 mm 2.7 mm Cable release 0,75 N +/- 0,15 N 0,4 N/mm 0.3 % 0.1 µm 0.2 µm 2 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.09 µm/°C Mahr LVDT 0.5 µm
5313049 Inductive probe
1304 K
5313049 -1 mm 1 mm 1304 K 1.1 mm 1.1 mm 0,75 N +/- 0,15 N 0,15 N/mm 1 % 0.15 µm 0.2 µm 4 µm IP 62 2.5 m 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Mahr LVDT 1 µm
5323010 Inductive probe
P2004 M
5323010 -2 mm 2 mm P2004 M -2.2 mm 0 mm 4.4 mm Standard model 0,75 N +/- 0,15 N 0,2 N/mm 0.3 % 0.1 µm 0.5 µm 1.5 µm 3 µm IP 64 2.5 m 20 °C 10 °C 40 °C 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Mahr VLDT 0.4 µm 2.2 mm
5323020 Inductive probe
P2004 MA
5323020 -2 mm 2 mm P2004 MA -2.2 mm 0 mm 4.4 mm Vacuum lifter 0,75 N +/- 0,15 N 0,2 N/mm 0.3 % 0.1 µm 0.5 µm 1.5 µm 3 µm IP 64 2.5 m 20 °C 10 °C 40 °C 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Mahr VLDT 0.4 µm 2.2 mm
5323030 Inductive probe
P2004 MB
5323030 -2 mm 2 mm P2004 MB -2.2 mm 0 mm 4.4 mm Compressed air retraction (max. 1 bar) Depending upon air pressure 0.3 % 0.1 µm 0.5 µm 1.5 µm 3 µm IP 64 2.5 m 20 °C 10 °C 40 °C 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Mahr VLDT 0.4 µm 2.2 mm
5324070 Inductive probe
P2104 MA
5324070 -2 mm 2 mm P2104 MA -2.2 mm 0 mm 10.4 mm Vacuum lifter 0,75 N +/- 0,15 N 0,1 N/mm 0.3 % 0.2 µm 1 µm 2 µm 4 µm IP 64 2.5 m 20 °C 10 °C 40 °C 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Mahr VLDT 0.5 µm 8.4 mm
5324080 Inductive probe
P2104 MB
5324080 -2 mm 2 mm P2104 MB -2.2 mm 0 mm 10.4 mm Compressed air retraction (max. 1 bar) Depending upon air pressure 0.3 % 0.2 µm 1 µm 2 µm 4 µm IP 64 2.5 m 20 °C 10 °C 40 °C 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Mahr VLDT 0.5 µm 8.4 mm
5323040 Inductive probe
P2001 M
5323040 -0.5 mm 0.5 mm P2001 M 0,75 N +/- 0,15 N 0,1 N/mm 0.3 % 0.15 µm 0.2 µm IP 40 2.5 m 20 °C 10 °C 40 °C 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Mahr VLDT 1.5 µm 0.6 µm
4400180 Inductive probe
P1300 MA
4400180 -2 mm 2 mm P1300 MA -2.2 mm 0 mm 4.4 mm Vacuum lifter (standard option) 0,75 N +/- 0,15 N 0,3 N/mm 0.3 % 0.1 µm 0.5 µm 1.5 µm 3 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Mahr half-bridge 0.4 µm 2.2 mm
4400181 Inductive probe
P1300 MB
4400181 -2 mm 2 mm P1300 MB -2.2 mm 0 mm 4.4 mm Pneumatic jacking (max. 1 bar) Depending upon air pressure 0.3 % 0.1 µm 0.5 µm 1.5 µm 3 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Mahr half-bridge 0.4 µm 2.2 mm
5313180 Inductive probe
1318
5313180 -0.3 mm 1 mm 1318 -0.37 mm 1.6 mm 0,25 N +/- 0,05 N 0,04 N/mm 0.5 % 0.03 µm 0.5 µm IP 50 1.5 m 10 °C 40 °C -10 °C 80 °C Mahr LVDT 0.9 µm
To top