Millimar | Module for inductive probes

Millimar N 1704 T

Art. no. 5331141

Module for inductive probes Millimar N 1704 T

  • Flexibly configurable RS485 bus modules
  • Powerful connection modules for evaluating measuring sensors (inductive/pneumatic)
  • Synchronous data retrieval from multiple connected measuring probes
  • Connection of N 1700 modules via USB interface to the smart, universal evaluation and configuration software Millimar Cockpit
  • Connects all compatible probe types using one module
  • Flexible and modular product combinations to tackle customer-specific measuring tasks
  • Maximum theoretical bus data rate of 4189 values/s (depending on the number of connected channels)

instruction manual

Millimar N 1704 T
  • Flexibly configurable RS485 bus modules
  • Powerful connection modules for evaluating measuring sensors (inductive/pneumatic)
  • Synchronous data retrieval from multiple connected measuring probes
  • Connection of N 1700 modules via USB interface to the smart, universal evaluation and configuration software Millimar Cockpit
  • Connects all compatible measuring probe types using one module
  • Flexible and modular product combinations to tackle customer-specific measuring tasks
  • Maximum theoretical bus data rate of 4189 values/s (depending on the number of connected channels)
Millimar N 1704 T | Art. no. 5331141
Product type N 1704 T
Resolution µm 0,1
Measuring range, inductive probe ± 5000 , ± 2000 , ± 1000 , ± 500
Probe inputs
Compatibility Tesa
Configuration Millimar Cockpit Software
Data transmission rate 4189 Hz
Error limit 0,3 % (min. 0,2 μm)
Data interface RS-485
Energy supply + 5V from N 1700 Bus
Current consumption 150 mA
IP protection category IP 42
Working temperature MIN 10 °C
Working temperature MAX 35 °C
Operating temperature Min 0 °C
Operating temperature Max 40 °C
Temperature coefficient 0.02 µm/°C
Product weight 0.394 kg

Variety of applications for our products

Smart and flexible combination of measuring modules and software for solving customer-specific measuring tasks.
Label Art. no. Measuring range from Measuring range up to Product type Distance of lower stop MIN Distance of lower stop MAX Distance of upper stop MIN Distance of upper stop MAX Lifter / retraction Measuring force (N) Increase in measuring force Sensitivity deviation Repeatability fw (µm) Measuring value hysteresis fu (µm) Linearity deviation within +/- 0,5 mm Linearity deviation within +/- 1,0 mm IP protection category Cable length Operating temperature Min Operating temperature Max Storage temperature MIN Storage temperature MAX Temperature coefficient Compatibility Linearity deviation within +/- 2,0 mm Working temperature Working temperature MIN Working temperature MAX Linearity deviation within +/- 5,0 mm
4400190 Inductive probe
P1300 TA
4400190 -2 mm 2 mm P1300 TA -2.2 mm 0 mm 2.2 mm 4.4 mm Vacuum lifter (standard option) 0,75 N +/- 0,15 N 0,3 N/mm 0.3 % 0.1 µm 0.5 µm 1 µm 3 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Tesa
4400192 Inductive probe
P1300 TA without cable
4400192 -2 mm 2 mm P1300 TA without cable -2.2 mm 0 mm 2.2 mm 4.4 mm Vacuum lifter (standard option) 0,75 N +/- 0,15 N 0,3 N/mm 0.3 % 0.1 µm 0.5 µm 1 µm 3 µm IP 64 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Tesa
4400191 Inductive probe
P1300 TB
4400191 -2 mm 2 mm P1300 TB -2.2 mm 0 mm 2.2 mm 4.4 mm Compressed air retraction (max. 1 bar) Depending upon air pressure 0.3 % 0.1 µm 0.5 µm 1 µm 3 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Tesa
4400193 Inductive probe
P1300 TB without cable
4400193 -2 mm 2 mm P1300 TB without cable -2.2 mm 0 mm 2.2 mm 4.4 mm Compressed air retraction (max. 1 bar) Depending upon air pressure 0.3 % 0.1 µm 0.5 µm 1 µm 3 µm IP 64 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Tesa
5323011 Inductive probe
P2004 T
5323011 -2 mm 2 mm P2004 T -2.2 mm 0 mm 2.2 mm 4.4 mm Standard model 0,75 N +/- 0,15 N 0,2 N/mm 0.3 % 0.1 µm 0.5 µm 0.4 µm 1.5 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Tesa 3 µm 20 °C 10 °C 40 °C
5323021 Inductive probe
P2004 TA
5323021 -2 mm 2 mm P2004 TA -2.2 mm 0 mm 2.2 mm 4.4 mm Vacuum lifter 0,75 N +/- 0,15 N 0,2 N/mm 0.3 % 0.1 µm 0.5 µm 0.4 µm 1.5 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Tesa 3 µm 20 °C 10 °C 40 °C
5323031 Inductive probe
P2004 TB
5323031 -2 mm 2 mm P2004 TB -2.2 mm 0 mm 2.2 mm 4.4 mm Compressed air retraction (max. 1 bar) Depending upon air pressure 0.3 % 0.1 µm 0.5 µm 0.4 µm 1.5 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Tesa 3 µm 20 °C 10 °C 40 °C
5324021 Inductive probe
P2010 TA
5324021 -5 mm 5 mm P2010 TA -5.3 mm 5.3 mm Vacuum lifter 0,75 N +/- 0,15 N 0,1 N/mm 0.3 % 0.2 µm 1 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Tesa 4 µm 20 °C 10 °C 40 °C 20 µm
5324031 Inductive probe
P2010 TB
5324031 -5 mm 5 mm P2010 TB -5.3 mm 5.3 mm Compressed air retraction (max. 1 bar) Depending upon air pressure 0.3 % 0.2 µm 1 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Tesa 4 µm 20 °C 10 °C 40 °C 20 µm
5324071 Inductive probe
P2104 TA
5324071 -2 mm 2 mm P2104 TA -2.2 mm 0 mm 8.4 mm 10.4 mm Vacuum lifter 0,75 N +/- 0,15 N 0,1 N/mm 0.3 % 0.2 µm 1 µm 0.5 µm 2 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Tesa 4 µm 20 °C 10 °C 40 °C
5324081 Inductive probe
P2104 TB
5324081 -2 mm 2 mm P2104 TB -2.2 mm 0 mm 8.4 mm 10.4 mm Compressed air retraction (max. 1 bar) Depending upon air pressure 0.3 % 0.2 µm 1 µm 0.5 µm 2 µm IP 64 2.5 m 10 °C 40 °C -10 °C 80 °C 0.15 µm/°C Tesa 4 µm 20 °C 10 °C 40 °C
To top