Millimar |
Module for inductive probes
Millimar N 1702 T
Art. no. 5331121
Module for inductive probes Millimar N 1702 T
- Flexibly configurable RS485 bus modules
- Powerful connection modules for evaluating measuring sensors (inductive/pneumatic)
- Synchronous data retrieval from multiple connected measuring probes
- Connection of N 1700 modules via USB interface to the smart, universal evaluation and configuration software Millimar Cockpit
- Connects all compatible probe types using one module
- Flexible and modular product combinations to tackle customer-specific measuring tasks
- Maximum theoretical bus data rate of 4189 values/s (depending on the number of connected channels)
instruction manual
- Flexibly configurable RS485 bus modules
- Powerful connection modules for evaluating measuring sensors (inductive/pneumatic)
- Synchronous data retrieval from multiple connected measuring probes
- Connection of N 1700 modules via USB interface to the smart, universal evaluation and configuration software Millimar Cockpit
- Connects all compatible measuring probe types using one module
- Flexible and modular product combinations to tackle customer-specific measuring tasks
- Maximum theoretical bus data rate of 4189 values/s (depending on the number of connected channels)
Millimar N 1702 T | Art. no. 5331121 | |
---|---|
Product type | N 1702 T |
Resolution µm | 0,1 |
Measuring range, inductive probe | ± 5000 , ± 2000 , ± 1000 , ± 500 |
Probe inputs | 2 |
Compatibility | Tesa |
Configuration | Millimar Cockpit Software |
Data transmission rate | 4189 Hz |
Error limit | 0,3 % (min. 0,2 μm) |
Data interface | RS-485 |
Energy supply | + 5V from N 1700 Bus |
Current consumption | 95 mA |
IP protection category | IP 42 |
Working temperature MIN | 10 °C |
Working temperature MAX | 35 °C |
Operating temperature Min | 0 °C |
Operating temperature Max | 40 °C |
Temperature coefficient | 0.02 µm/°C |
Product weight | 0.27 kg |
Variety of applications for our products
Smart and flexible combination of measuring modules and software for solving customer-specific measuring tasks.
Label | Art. no. | Measuring range from | Measuring range up to | Product type | Distance of lower stop MIN | Distance of lower stop MAX | Distance of upper stop MIN | Distance of upper stop MAX | Lifter / retraction | Measuring force (N) | Increase in measuring force | Sensitivity deviation | Repeatability fw (µm) | Measuring value hysteresis fu (µm) | Linearity deviation within +/- 0,5 mm | Linearity deviation within +/- 1,0 mm | IP protection category | Cable length | Operating temperature Min | Operating temperature Max | Storage temperature MIN | Storage temperature MAX | Temperature coefficient | Compatibility | Linearity deviation within +/- 2,0 mm | Working temperature | Working temperature MIN | Working temperature MAX | Linearity deviation within +/- 5,0 mm | |
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Inductive probe P1300 TA |
4400190 | -2 mm | 2 mm | P1300 TA | -2.2 mm | 0 mm | 2.2 mm | 4.4 mm | Vacuum lifter (standard option) | 0,75 N +/- 0,15 N | 0,3 N/mm | 0.3 % | 0.1 µm | 0.5 µm | 1 µm | 3 µm | IP 64 | 2.5 m | 10 °C | 40 °C | -10 °C | 80 °C | 0.15 µm/°C | Tesa | ||||||
Inductive probe P1300 TB |
4400191 | -2 mm | 2 mm | P1300 TB | -2.2 mm | 0 mm | 2.2 mm | 4.4 mm | Compressed air retraction (max. 1 bar) | Depending upon air pressure | 0.3 % | 0.1 µm | 0.5 µm | 1 µm | 3 µm | IP 64 | 2.5 m | 10 °C | 40 °C | -10 °C | 80 °C | 0.15 µm/°C | Tesa | |||||||
Inductive probe P2004 T |
5323011 | -2 mm | 2 mm | P2004 T | -2.2 mm | 0 mm | 2.2 mm | 4.4 mm | Standard model | 0,75 N +/- 0,15 N | 0,2 N/mm | 0.3 % | 0.1 µm | 0.5 µm | 0.4 µm | 1.5 µm | IP 64 | 2.5 m | 10 °C | 40 °C | -10 °C | 80 °C | 0.15 µm/°C | Tesa | 3 µm | 20 °C | 10 °C | 40 °C | ||
Inductive probe P2004 TA |
5323021 | -2 mm | 2 mm | P2004 TA | -2.2 mm | 0 mm | 2.2 mm | 4.4 mm | Vacuum lifter | 0,75 N +/- 0,15 N | 0,2 N/mm | 0.3 % | 0.1 µm | 0.5 µm | 0.4 µm | 1.5 µm | IP 64 | 2.5 m | 10 °C | 40 °C | -10 °C | 80 °C | 0.15 µm/°C | Tesa | 3 µm | 20 °C | 10 °C | 40 °C | ||
Inductive probe P2004 TB |
5323031 | -2 mm | 2 mm | P2004 TB | -2.2 mm | 0 mm | 2.2 mm | 4.4 mm | Compressed air retraction (max. 1 bar) | Depending upon air pressure | 0.3 % | 0.1 µm | 0.5 µm | 0.4 µm | 1.5 µm | IP 64 | 2.5 m | 10 °C | 40 °C | -10 °C | 80 °C | 0.15 µm/°C | Tesa | 3 µm | 20 °C | 10 °C | 40 °C | |||
Inductive probe P2010 TA |
5324021 | -5 mm | 5 mm | P2010 TA | -5.3 mm | 5.3 mm | Vacuum lifter | 0,75 N +/- 0,15 N | 0,1 N/mm | 0.3 % | 0.2 µm | 1 µm | IP 64 | 2.5 m | 10 °C | 40 °C | -10 °C | 80 °C | 0.15 µm/°C | Tesa | 4 µm | 20 °C | 10 °C | 40 °C | 20 µm | |||||
Inductive probe P2010 TB |
5324031 | -5 mm | 5 mm | P2010 TB | -5.3 mm | 5.3 mm | Compressed air retraction (max. 1 bar) | Depending upon air pressure | 0.3 % | 0.2 µm | 1 µm | IP 64 | 2.5 m | 10 °C | 40 °C | -10 °C | 80 °C | 0.15 µm/°C | Tesa | 4 µm | 20 °C | 10 °C | 40 °C | 20 µm | ||||||
Inductive probe P2104 TA |
5324071 | -2 mm | 2 mm | P2104 TA | -2.2 mm | 0 mm | 8.4 mm | 10.4 mm | Vacuum lifter | 0,75 N +/- 0,15 N | 0,1 N/mm | 0.3 % | 0.2 µm | 1 µm | 0.5 µm | 2 µm | IP 64 | 2.5 m | 10 °C | 40 °C | -10 °C | 80 °C | 0.15 µm/°C | Tesa | 4 µm | 20 °C | 10 °C | 40 °C | ||
Inductive probe P2104 TB |
5324081 | -2 mm | 2 mm | P2104 TB | -2.2 mm | 0 mm | 8.4 mm | 10.4 mm | Compressed air retraction (max. 1 bar) | Depending upon air pressure | 0.3 % | 0.2 µm | 1 µm | 0.5 µm | 2 µm | IP 64 | 2.5 m | 10 °C | 40 °C | -10 °C | 80 °C | 0.15 µm/°C | Tesa | 4 µm | 20 °C | 10 °C | 40 °C |
Products |
Variants and options